| Type: |
microprocessor-controlled, programmable syringe pump; infusion/withdrawal and valve control |
| Program: |
up to 99 steps of speed and time |
| Time resolution: |
0 to 999 minutes in 1 minute steps or 0 to 99.9 minutes in 0.1 minute steps |
| Maximum program length: |
1650 hours |
| Accuracy: |
± 1 % |
| Reproducibility: |
± 0.2 % (electronically) |
| Syringes: |
glass, plastic, metal syringes; with volumes from 0.5 µl to over 150 ml |
| Flow rate range: |
minimum 0.4 nl/min. with a 0.5 µl syringe; maximum 110 ml/min. (6.6 l/hr) with a 150 ml syringe |
| Non-volatile memory: |
storage of all settings |
| Maximum force: |
VIT-FIT syringe pump: 30 kg (reducible to 8 kg by a switch) VIT-FIT HP (high-pressure) syringe pump: 60 kg (reducible to 16 kg by a switch) |
| Motor: |
microprocessor controlled, long life brushless BLDC motor with neodymium magnets |
| Transmission: |
efficient force transmission by a ball screw with highest mechanical load capacity of 12 000 N |
| Pusher travel: |
120 mm |
| Pusher travel rate: |
minimum 0.08 mm/min; maximum 80 mm/min |
| Speed range: |
000 to 999 |
| Interface: |
RS-485 or RS-232 (option) |
| Voltage: |
95 to 240 V / 50–60 Hz AC (input) plug integrated switching power supply; DC 12 V / 24 W (output) possible field operation on 12 V accumulator |
| Dimensions: |
12.5 × 26.5 × 13 cm (H × W × D) |
| Weight: |
3.5 kg |